Message: Return type of CI_Session_files_driver::open($save_path, $name) should either be compatible with SessionHandlerInterface::open(string $path, string $name): bool, or the #[\ReturnTypeWillChange] attribute should be used to temporarily suppress the notice
Message: Return type of CI_Session_files_driver::close() should either be compatible with SessionHandlerInterface::close(): bool, or the #[\ReturnTypeWillChange] attribute should be used to temporarily suppress the notice
Message: Return type of CI_Session_files_driver::read($session_id) should either be compatible with SessionHandlerInterface::read(string $id): string|false, or the #[\ReturnTypeWillChange] attribute should be used to temporarily suppress the notice
Message: Return type of CI_Session_files_driver::write($session_id, $session_data) should either be compatible with SessionHandlerInterface::write(string $id, string $data): bool, or the #[\ReturnTypeWillChange] attribute should be used to temporarily suppress the notice
Message: Return type of CI_Session_files_driver::destroy($session_id) should either be compatible with SessionHandlerInterface::destroy(string $id): bool, or the #[\ReturnTypeWillChange] attribute should be used to temporarily suppress the notice
Message: Return type of CI_Session_files_driver::gc($maxlifetime) should either be compatible with SessionHandlerInterface::gc(int $max_lifetime): int|false, or the #[\ReturnTypeWillChange] attribute should be used to temporarily suppress the notice
Syringe pump KSP–A100 is a light weight, portable device designed for low-volume infusion. Equipped with LCD screen to display parameters like injection state, volume, mode, speed, syringe size, alarms, dosage, occlusion status, etc. for enhanced clinical efficiency. Inbuilt software based control system delivers higher levels of precision, safety and allows remote control mechanism to improve standardization, accuracy analysis of infusion data and reduce free–flow risks.
Specifications :
KVO Rate
0.1 – 5ml/h (steps : 0.1ml/h)
Injection Compliance
Adaptable for 5 syringe sizes
Preset Volume Range
0.1 ml – 1999.9 ml
Occlusion System
3-level alarming system
Mechanical Precision
≤ ±2 %
Injection Rate Range
50 ml Syringe
30 ml Syringe
20 ml Syringe
10 ml Syringe
5 ml Syringe
0.1 ml – 999.9 ml/h (step: 0.1 ml/h) 1000 ml – 1800 ml/h (step: 1 ml/h)
0.1 ml – 900 ml/h (step: 0.1 ml/h)
0.1 ml – 600 ml/h (step: 0.1 ml/h)
0.1 ml – 300 ml/h (step: 0.1 ml/h)
0.1 ml – 150 ml/h (step: 0.1 ml/h)
Purge Rate
50 ml Syringe
30 ml Syringe
20 ml Syringe
10 ml Syringe
5 ml Syringe
1800 ml/h (Bolus rate: 1200 ml/h)
900 ml/h (Bolus rate: 600 ml/h)
600 ml/h (Bolus rate: 400 ml/h)
300 ml/h (Bolus rate: 200 ml/h)
150 ml/h (Bolus rate: 100 ml/h)
Occlusion Alarm Threshold
High
Medium
Low
800 mmHg ± 200 mmHg (106.7 kPa ± 26.7 kPa)
500 mmHg ± 100 mmHg (66.7 kPa ± 13.3 kPa)
300 mmHg ± 100 mmHg (40.7 kPa ± 13.3 kPa)
Fuse
F1AL/250V (2 nos.)
Power Supply
100 V – 220 V, 50 Hz
Battery
Built–in Rechargeable Li–ion battery (11.1V) Capacity ≥ 2000mAh ; 10 hours of charging Works at 5ml/h after 4 hours of charging
Alarm
Near Injection finished, Preset dose injection finished, Occlusion,
Syringe off, Syringe dislocated, Low Battery, On AC Supply.
Safety
Class I Type BF
Operation
Ambient Temperature
Relative Humidity
Atmospheric Pressure
5°C to 40°C
20% – 90%
86 kPa – 106 kPa
Storage
Ambient Temperature
Relative Humidity
Atmospheric Pressure
–30°C to 55°C
≤ 95%
50 kPa – 106 kPa
Dimensions
282 x 210 x 148 mm
Weight
2 Kgs
Features :
4.3” color segment LCD
Multi injection modes – Volume/Time/Body
Adjustable speed, time, volume and dosage
Audio–visual alarms for critical events
Applications :
Used in general ward, ICU or post-operative care facilities in hospitals to gradually administer intravenous fluid drug dosage in controlled amounts at a predetermined rate and speed through syringe attached with a catheter for consistent and effective infusion therapy.
Meditronics